<7. Setting Parameters>

7-16

IM 01C50E01-01EN

Input

Output

GAIN

F0705.ai

ZERO

Figure 7.3

Trim function images

IMPORTANT

Setting the sensor type automatically resets the

zero-gain adjustment value to the factory setting.

2) Clearing the Adjustment Value

The adjustment value set with input adjustment

can be cleared and reset to the factory setting.

■ Procedure to call up the adjustment value reset

parameter (Cal Point Clear)

AI1/AI2 block: Cal Point Clear
Set the Cal Point Clear parameter to Clear.

7.3.14 Switching to the Deep Sleep Mode

When the instrument will not be used for a long

time, switch the instrument to the deep sleep mode

to conserve battery power. To switch to deep sleep

mode, follow the procedure below.

■ Procedure to call up the switch-to-deep-sleep

parameter (Special Cmd)

TRANSDUCER block : Special Cmd
Set deep sleep mode (Standby) to the Special

Cmd parameter for the TRANSDUCER block.

To start from deep sleep mode, remove and

insert the battery pack.

CAUTION

After setting the deep sleep mode by infrared

device configuration tool, keep the infrared port

of device away from any other infrared signals.

NOTE

• This procedure to switch to deep sleep mode

can only be use for wireless communication.

• Transmitter becomes the stop state after

setting deep sleep mode and cannot reply

any request from Communication Tool.

• For this reason, there is the case that an

error is display on Communication Tool.

• To wake up from deep sleep mode, please

pull battery pack and wait more than 30

seconds before attaching battery pack.

7.3.15 Switching to the Silence Mode

This is a function to pause the instrument when

it cannot join the field wireless network after a

specified time has elapsed. This function is effective

in conserving battery power when, for example, the

installation of the field wireless integrated gateway

is delayed compared to that of field wireless

devices. The default value is 28800 seconds

(about 8 hours). Thereafter, a cycle of a one-hour

pause and six-minute search is repeated until the

instrument can join the field wireless network.

• Procedure to call up the switch-to-silence

parameter (Radio Silence)

UAPMO block : Radio Silence
Set 0 to 2

31

seconds for the Radio Silence

parameter of the UAPMO block. If 0 is set, the

Radio Silence parameter is invalid. To start from

the silence mode, either remove and insert the

battery pack, or use the configuration tool.