7. Startup

7.11.2

Checking Calibration Contact Output

The calibration contacts are used for solenoid valve drive signals for the ZR40H
Automatic Calibration Unit. When using the ZR40H Automatic Calibration Unit, use the
calibration contact output to check that the wiring connections have been properly
completed and check equipment operation.

(1) Referring to Section 7.11.1, display the contact check display.
(2) Select the Calibration contacts to display the panel display as Figure 7.15.3 shows.
(3) Open the zero-gas contact and the span-gas contact. This will help check the auto-

matic calibration unit and wiring connections.

Zero gas contact : Open

Span gas contact : Open

Enter

Calibration contacts

F7.15.3E.EPS

Figure 7.15.3 Calibration Contact Check Display

Note

“Open” and “Closed” displayed on the Calibration contact display indicate actions of
drive contacts and are opposite to the valve open and close actions. If Open is displayed
on the Calibration contact display, no calibration gas flows. If Closed is displayed on
that display, calibration gas flows.

7.11.3

Checking Input Contacts

(1) Referring to Section 7.11.1, display the contact check display.
(2) Display the input contact check display as Figure 7.15.4 shows. The “open” or

“closed” input contact in the display shows the current contact input terminal status
and the display changes according to the contact status. Using this enables you to
check that the wiring connections have been properly completed.

Input contact 1 : Open

Input contact 2 : Open

Enter

Input contacts

F7.15.4E.EPS

Figure 7.15.4 Input Contact Check Display