IM 11M13A01-02E

2-28

2.7.3

Check Valve (Part No. K9292DN or K9292DS)

This valve is mounted on the calibration gas line (directly connected to the detector).
This is applied to a system based on the (System 2 and 3) system configuration .

This valve prevents the process gas from entering the calibration gas line. Although it
functions as a stop valve, operation is easier than a stop valve as it does not require
opening/closing at each calibration.

Screw a check valve, instead of a stop valve into the calibration gas inlet of the detector.

Standard Specifications
Connection:

Rc1/4 or 1/4NPT

Material:

SUS304 (JIS)

Pressure:

70 kPa G or more and 350 kPa G or less

Weight:

Approx. 40 g

Joint: Rc 1/4, Material: SUS304 (JIS)

Joint: 1/4 NPT, Material: SUS304 (JIS)

T2.10E.EPS

Part No.

K9292DN

K9292DS

Description

K9292DN : Rc1/4 (A part), R1/4 (B part)
K9292DS : 1/4NPT female (A part),1/4NPT male (B part)

F2.11E.EPS

Approx.19

Approx.54

A

B

Unit : mm