General Specifications

Standard Specifications
Measured Object:

Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases. May not be applicable corrosive gas
such as ammonia and chlorine.)
The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.

Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O

2

Output Signal:

4 to 20 mA DC (maximum load resistance 550

⍀)

Measurement Range:

Any setting in the range of 0 to 5 through 0 to 100 vol% O

2

(in

1 vol% O

2

), or partial range

Digital Communication (HART): 250 to 550

⍀, depending on number of field devices

connected to the loop (multi-drop mode).

Note:

HART is a registered trademark of the HART Communication Foundation.

Display Range:

0 to 100 vol% O

2

Warm-up Time:

Approx. 20 min.

Repeatability:

؎0.5% Maximum value of range setting. (range up to 0 to 25 vol%
O

2

)

؎1% Maximum value of range setting. (range from 0 to 25 vol%
O

2

up to 0 to 100 vol% O

2

)

Linearity:

(Excluding standard gas tolerance)
(Use oxygen of known concentration (in the measuring range) as
the zero and span calibration gas.)
؎1% Maximum value of range setting ; from 0 to 5 vol% O

2

to 0

to 25 vol% O

2

range

(Sample gas pressure: within

؎4.9 kPa)

؎3% Maximum value of range setting ; from 0 to 25 vol%O

2

to 0

to 50 vol% O

2

range

(Sample gas pressure: within

؎0.49 kPa)

؎5% Maximum value of range setting ; from 0 to 50 vol% O

2

to 0

to 100 vol% O

2

range (Sample gas pressure: within

؎0.49 kPa)

Drift:

(Excluding the first two weeks in use)
Both zero and span

؎2% Maximum value of range setting/month

Response Time :

Response of 90% within 5 seconds. (Measured from when gas is
introduced from calibration gas inlet and analog output start
changing.)