6 contact output wiring, 1 cable specifications, 2 wiring procedure – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 91

Advertising
background image

IM 11M13A01-02E

5-12

5.6

Contact Output Wiring

Contact outputs 1 to 3 can be freely assigned to "low limit alarm", "high limit alarm",
etc. user selectable, but the assignment of contact output 4 is fixed ("error output"). And
the action (contact closed on error output) also cannot be changed.

When using these contact outputs, install the wiring as follows:

#1 Output

#2 Output

#3 Output

#4 Output

DO-1

DO-1

DO-2

DO-2

DO-3

DO-3

Terminal box

Annunciator or the like

F5.10E.EPS

Converter

DO-4

DO-4

Figure 5.10 Contact output wiring

5.6.1

Cable Specifications

Number of wire in cable varies depending on the number of contact used.

5.6.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each wire in the cable should

be terminated in the corresponding crimp-on terminal.

(2) The capacities of the contact output relay are 30 V DC 3 A, 250 V AC 3 A. Connect

a load (e.g. pilot lamp and annunciator) within these limits.