2 default values, Caution, Calibration timing – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 149

Advertising
background image

IM 11M13A01-02E

9-8

(1) Select the “Calibration timing” display. A panel display as shown in Figure 9.7

appears.

(2) Select each item for the calibration to display the numeric-data entry display. Enter

the desired numeric values for the calibration.

Enter

Hold time: 1 0 min 0 0 s

Csl time: 1 0 min 0 0 s

lnterval: 3 0 d 0 0 h

Start date: 0 1 / 0 1 / 0 0

Start time: 0 0 : 0 0

Calibration timing

F9.7E.EPS

Figure 9.7 “Calibration Timing” Display

CAUTION

When setting calibration timing requirements, bear the following precautions in mind:

(1) If the calibration interval is shorter than the sum of stabilization time plus calibration

time, the second calibration start time will conflict with the first calibration. In such a
case, the second calibration will not be conducted. (When both zero and span
calibrations are to be performed, the calibration time is double that required for a
single (zero or span) calibration.)

(2) For the same reason, if the calibration start time conflicts with manual calibration or

semi-automatic calibration, the current calibration will not be conducted.

(3) If the calibration time conflicts with maintenance service or blowback operations,

calibration will start after completing the maintenance service or blowback opera-
tions (see Section 8.2.1, earlier in this manual).

(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration

will be conducted; a second or later calibration will not be conducted.

(5) If a past date is set to the calibration start date, no calibration will be conducted.

9.2.2

Default Values

When the analyzer is delivered, or if data are initialized, the calibration settings are by
default, as shown in Table 9.1.

Table 9.1 Default Settings for Calibration

Item

Default Setting

Calibration mode

Manual

Calibration procedure

Span - zero

Zero-gas (oxygen)

1.00%

concentration

Span-gas (oxygen) concentration

21.00%

Output hold (stabilization) time

10 minutes, 00 seconds

Calibration time

10 minutes, 00 seconds

Calibration interval

30 days, 00 hours

Start date

01 / 01 / 00

Start time

00:00

Manual
Calibration

Semi-automatic
Calibration

Automatic
Calibration

T9.1E.EPS