General specifications, 1. specifications – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 14

Advertising
background image

IM 12D08B02-01E

Specifications 2-1

2-1. Specifications
A. Input specifications

:

Two or four electrodes measurement with

square wave excitation. Cell constants from

0.008 to 50 cm

-1

.

B. Detection method

:

Frequency, read-pulse position and reference

voltage are dynamically optimized.

C. Input ranges

- Conductivity :

Minimum : 0 μS/cm

Maximum : 200 mS x (Cell constant)

(overrange 1999 mS / cm).

- Resistivity :

Minimum : 0.005 kΩ/ (Cell constant)

Maximum : 999 MΩ x cm

- Temperature

Pt1000

: -20 to +250 °C (0 to 500 °F)

Pt100 and Ni100 : -20 to +200 °C (0 to 400 °F)

8K55 NTC : -10 to +120 °C (10 to 250 °F)

PB36 NTC : -20 to +120 °C (0 to 250 °F)

D. Output Span

- Conductivity : - min 0.01μS/cm

: - max. 1999 mS/cm. (max 90%

zero suppression)

- Resistivity : - min 0.001kΩxcm

: - max. 999 MΩ x cm. (max 90%

zero suppression)

E. Transmission Signal

: Isolated output of 4-20 mA DC

Burn up (21 mA) or Burn down

(3.6 mA when HART® or distributor comm. is

non-used, 3.9 mA when HART® or distributor

comm. is used) or pulse of 21 mA to signal

failure.

F. Temperature compensation

:

Automatic, for temperature ranges mentioned

under C (inputs).

- Reference temp.

: programmable from 0 to 100 °C or 30 to 210 °F

(default 25 °C).

G. Compensation algorithm

-NaCl

:

According IEC 60746-3 NaCl tables

(default).

-T.C.

:

Two independent user program-

mable temperature coefficients,

from -0.0% to 3.5% per °C (°F)

by adjustment or calibration.

- Matrix

: Conductivity function of concentra-

tion and temperature. Choice out of

mA

mA

mA

mA

5 preprogrammed matrixes and a 25-

point user-programmable matrix.

H. Logbook

:

Software record of important events and diag-

nostic data. Available through HART® link,

with diagnostic information available in the

display.

I. Display

:

Custom liquid crystal display, with a main dis-

play of 31/2 digits 12.5 mm high. Message dis-

play of 6 alpha numeric characters, 7 mm high.

Warning flags and units (mS/cm, kΩ·cm, μS/

cm and MΩ·cm) as appropriate.

J. Power supply

:

Nominal 24 volt DC loop powered system.

SC202G

; 17 to 40 volts, see Fig.2-1

SC202S

: 17 to 31.5 volts, see Fig.2-2

Maximum load resistance

For the SC202G, see Fig. 1

200Ω or less with the PH201G

50Ω or less with the SDBT

For the SC202S, see Fig. 2-2

1000

1150

800

600

400

200

150

0

10

20

18

17

30

40

0

Voltage (V)

Load Resistance (

Ω

)

Possible

F06.EPS

Fig.2-1 Supply voltage/ load diagram for the SC202G

0

200

400

600

800

12

16

17

20

24

28

32

425

775

Voltage (V)

Load Resistance (

Ω

)

31.5 V

Possible

Fig.2-2 Supply voltage/ load diagram for the SC202S

K. Input isolation

:

1000

VDC

L. Weight

Body weight

:

approx. 1.6 kg

Mounting brackets weight: approx. 0.7 kg.

2. GENERAL SPECIFICATIONS

Advertising