2. calibration procedure, 2. calibration procedure -2 – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 67

Advertising
background image

IM 12D08B02-01E

6-2 Calibration

6-2. Calibration procedure

ENT

ENT

>

Select the flashing digit with the > key.

Increase its value by pressing the key

>

After the indication is stable, set the value

using the >, , ENT key.

Put the sensor in standard

solution. Press YES.

When the correct value is displayed,

press ENT to enter the change.

After briefing displaying WAIT,

the CAL.END message appears.

The calibration is now complete. Put the

sensor back in the process and press YES.

NO

YES

NO

YES

MODE

NO

YES

YES

ENT

NO

MODE

YES

MODE

MEASURE

DISPLAY

CAL

HOLD

Press the MODE key.

The legend CALIB

appears, and the YES/NO

key prompt flags flash.

ENT

The cell constant is automatically updated after the calibration and the new

value can be read on the display as described in section 4.5.

The calculation is as follows: Cell constant in /cm= (Conductivity of calibration solution in mS/cm) x

(Cell resistance in kOhm)

Comparing this calibrated cell constant with the initial nominal cell constant in service code 03

gives a good indication of the stability of the sensor. If the calibrated cell constant differs more than

20% from the nominal cell constant error E3 is displayed.