12-4. Supplement of parameter setting

12-4-1. Set cell constant (service code 03)

Code 3 *0.10xC First select a multiplying factor, and then set the constant in consideration of this fac-

tor. The position of the decimal point can be selected after the first digit has been set

(when the decimal point is flashing).

*How to enter the cell constant

(1) In the case that the only cell constant is mentioned on the text plate of the sensor (SC211G,

SC8SG, SC4AJ). How to enter the cell constant of 0.0195 /cm: Select *0.01xC on the message

display, and then enter the value of 1.950 on the main display.

(2) In the case that the deviation of a nominal cell constant (± X.X%) is mentioned on the text plate

of the sensor (SC210G).

When the nominal cell constant is 5 /cm and the deviation (CORR.% = -1.1) is mentioned:

The cell constant to be entered is calculated as follows:

5 + 5 x (-1.1/100) = 4.945

How to enter the cell constant of 4.945 /cm:

Select *10.0xC on the message display, and then enter the value of 0.495 (rounded to three

decimal places) on the main display. (The first digit in the constant setting only accepts 1 or 2.)

12-4-2. Temperature sensor (service code 10)

Code 10 *T.SENS Selection of the temperature compensation sensor. The default selection is the

PT1000 sensor, which gives excellent precision with the two wire connections used.

The other options give the flexibility to use a very wide range of other conductivity/

resistivity sensors. The temperature sensors for the applicable conductivity sensors

are as follows. According to the conductivity sensor used, select the appropriate

temperature sensor.

* SC210G

PB36NTC

* SC211G

Pt1000Ω

* SC8SG

Pt1000Ω

* SC4AJ

Pt1000Ω

Code Display

Function

Function detail

X Y Z Default values

Temperature measuring functions

10 *T.SENS Temperature

sensor Pt1000Ω

0 0 Pt1000Ω

Ni100Ω

1

PB36

(PB36NTC)

2

Pt100Ω

3

8k55

(8.55kΩNTC) 4

12-4-3. Automatic return (service code 50)

If no key is operated for 10 minutes in any mode other than measurement mode or after 10 minutes in

Hold status, Auto-Return (factory setting: On (1) in service code 50) will be activated to return the trans-

mitter to measurement mode. To disable Auto-Return, set the service code 50 to Off (0).

When stopping auto-return function, the transmitter doesn't automatically return to measurement mode. Take

care of returning measurement mode for re-measurement.

WARNING