2. wiring diagrams, Example of non-explosionproof system, B) sc4aj, sc8sg – Yokogawa EXA SC202 2-Wire Conductivity Transmitter/Analyzer User Manual

Page 84

Advertising
background image

IM 12D08B02-01E

2-2 Appendix

12-2. Wiring diagrams
1. Example of Non-Explosionproof System
(a) SC210G-A or SC210G-B

SC210G–A, SC210G–B

Conductivity sensor

SC202G

Conductivity transmitter

PH201G (Style B)

Dedicated distributor for EXA202

*1

*2

SDBT distributor

*2

11 Temper-

ature

sensor

Electrode

T1

11

T1

12

T2

12

T2

13

13

14

C1

14

C1

15

+

Output

(1 to 5V DC)

A(

+

)

B (

)

+

Output

(1 to 5V DC)

C

D

F

H

HOLD FAIL

Relay contacts

b a d c

+

G

C2

15

C2

16

Ground

16

Output

(1 to 5V DC)

+

Output

(1 to 5V DC)

+

1 (

+

)

2 (

)

A

B

F

H

F08.EPS

*3 (100Ω or less)

(b) SC4AJ, SC8SG

SC4AJ, SC8SG

Conductivity sensor

*1

11

11

12

12

13

13

14

14

15

15

16

16

F09.EPS

Temperature

sensor

Conductivity

sensor

*1 : This cable is specified by the additional code of an conductivity sensor.

*2 : Use a two-conductor shielded cable of OD 6 to 12mm.

The cable length is : Max. 2000m (also the minimum operating voltage of conductivity

transmitter must be obtained)

*3 : Conduct grounding without fail on the conductivity transmitter

(Grounding reistance : 100 ½ or less)

SC202G

Conductivity transmitter

2. Example of Intrinsically Safe Explosionproof System
(a) SC210G-A or SC210G-B

SC210G-A,SC210G-B

Conductivity sensor

T1
T2

C1
C2

T1
T2

C1
C2

*1

11
12
13
14
15
16

11
12
13
14
15
16

SC202S

Conductivity transmitter

Temper-

ature

sensor

Electrode

*2

Ground

to earth

+

-

G

Distributor

Safety Barrier

*1: This cable is specified by the additional code of an conductivity sensor.

*2: Use two-wire cable with OD (Outside Diameter) of 6 to 12 mm.

F007-1.eps

Output

(b) SC4AJ, SC8SG

SC4AJ, SC8SG

Conductivity sensor

*1

11

11

12

12

13

13

14

14

15

15

16

16

F09.EPS

Temperature

sensor

Conductivity

sensor

*1 : This cable is specified by the additional code of an conductivity sensor.

SC202S

Conductivity transmitter

Advertising