3-10 Appendix

2/3

QIS 12D08B02-61E

In this state, change the resistance value of the decade resistance box 1 as shown in Table 1.

The corresponding temperature indication must be within the range.

Table 1 Temperature Indication Check

Reference

Temperature

Resistance of

Resistance Box 1

Indication Range

–10 °C

960.9 :

–10

±0.3

°C

75 °C

1289.8 :

75

±0.3

°C

190 °C

1721.6 Ω

190

±0.3

°C

240 °C

1904.6 Ω

240

±0.3

°C

3.4 Conductivity Indication Check

Connect the instruments as shown in Figure 1, and set them as follows.

Decade resistance box 1:

100 Ω

Decade resistance box 2:

10 Ω

In this state, change the resistance value of the decade resistance box 2 as shown in Table 2.

The corresponding conductivity indication must be within the range.

Table 2 Conductivity Indication Check (Cell Constant : 0.1/cm)

Reference

Conductivity

Resistance of

Resistance Box 2

Indication Range

10 mS/cm

10 Ω

10

±0.05

mS/cm

1 mS/cm

100 :

1

±0.005

mS/cm

100 μS/cm

1 k:

100 ±0.5 μS/cm

10 μS/cm

10 k:

10

±0.05

μS/cm

1 μS/cm

100 k:

1

±0.005

μS/cm

3.5 Fieldbus

Communication

Functional

Check

Check for normal function using Fieldbus equipment specified by Yokogawa.