IM 12D08B02-01E

5-2 Parameter setting

5-2. Commissioning mode

5-2-1. Introduction

In order to obtain peak performance from the EXA SC202, you must set it up for each custom applica-

tion.

*OUTP

: mA output is set as default to 0-1 mS/cm or 0-19.99 MΩ·cm.

For enhanced resolution in more stable measuring processes, it may be desirable to

select for example 5-10 μS/cm range.

*HOLD

: The EXA SC202 transmitter has the ability to “HOLD” the output during maintenance

periods. This parameter should be set up to hold the last measured value, or a fixed

value to suit the process.

*TEMP.1, 2

: First and second temperature compensation types and values. (see also section 5-2-4)

* NaCl is the default compensation and is used for neutral salt solutions. Strong solu-

tions of salts are compensated, as are process waters and pure, and ultrapure water.

* TC temperature coefficient compensation uses a linear temperature compensation

factor. This can be set by calibration or configuration.

* Matrix compensation is an extremely effective way of compensation. Choose from

standard matrix tables, or configure your own to exactly suit your process.

*SERV

: This selection provides access to the service menu.

What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter

setting function. By following the simple YES/NO prompts and arrow keys, users can navigate through

the process of setting range, hold and service functions.