Installation and wiring, 1. installation and dimensions, 1-1. installation site – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 12: 1-2. mounting methods, Installation and wiring -1, 1. installation and dimensions -1, 1-1. installation site -1, 1-2. mounting methods -1, Warning

Advertising
background image

IM 12D08E02-01E

Installation and wiring 3-1

3. INSTALLATION AND WIRING

3-1. Installation and dimensions

3-1-1. Installation site

This instrument is a Class A product, and it is designed for use in the industrial environment. Please use

this instrument in the industrial environment only.

The EXA converter is weatherproof and can be installed inside or outside . It should, however, be

installed as close as possible to the sensor to avoid long cable runs between sensor and converter .

In any case, the total cable length should not exceed 60 meters (200 feet) . Select an installation site

where:

• Mechanical vibrations and shocks are negligible

• No relay/power switches are in the direct environment

• Access is possible to the cable glands (see figure 3-1)

• The converter is not mounted in direct sunlight or severe weather conditions

• Maintenance procedures are possible (avoiding corrosive environments)

The ambient temperature and humidity of the installation environment must be within the limits of the

instrument specifications. (See chapter 2).

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the EXA converter has universal mounting capabilities:

• Panel mounting using optional brackets

• Surface mounting on a plate (using bolts from the back)

• Wall mounting on a bracket (for example, on a solid wall)

• Pipe mounting using a bracket on a horizontal or vertical pipe (nominal pipe diameter 50A)

184

144

72

20

144

220

23

112

Four M6 screws, 8

(0.31)

deep

80

80

36

36

36

38

A

B

C

D

E

F

□ Hood (optional)
Option code : /H

Cable inlet port (21

(0.83)

dia. holes)

equivalent to DIN PG13.5 cable gland

A : For sensor cable

B : For sensor cable

C : For output signal

D : For contact output (S3 and S4)

E : For contact output (S1 and S2)

F : For power supply

Ground terminal

(M4 screw)

Weight: Approx. 2 kg

Adaptor for conduit work

(option code : /AFTG, /ANSI)

F17.ai

Approx. 55

Adaptor

G 1/2 female ( / AFTG)

1/2 NPT female ( / ANSI)

49

(7.24)

(8.66)

(3.15)

(3.15)

(1.93)

(2.17)

(5.67)

(5.67)

(2.83)

(0.79)

(0.91)

(4.41)

(1.42) (1.42)

(1.42)

(1.50)

Figure 3-1. Housing dimensions and layout of glands

WARNING

Advertising