5-20 Parameter setting

5-3-5. Contact outputs

Code 40 *S1, *S2

Process relays can be set for a variety of alarm and control function.

41, 42

*S3, *S4

and 43

Digit “X” sets the type of trigger:

Off means that the relay is not active

Low setpoint means that the relay is triggered by a decreasing measurement.

High setpoint means that the relay is triggered by an increasing measurement

“HOLD” active means that there is maintenance activity in progress so the

measurement is not live.

For *S4 There is the extra possibility to set up for “FAIL” indication.

Digit “Y” sets the control action:

Process alarm is a simple On/Off trip controlled by the high/low setpoint.

Proportional duty cycle control has a pulse width modulation for proportional

dosing with solenoid valves.

 Proportional frequency control is used for controlling electrically positioned

valves.

Temperature alarm is an On/Off trip on the measured temperature.

Digit “Z” sets the control parameter:

Alarm on main process

Control on main process

(Main process means conductivity/resistivity depending on the setting of

service code 01).

Code Display

Function

Function detail

X

Y

Z

Default values

Contacts
40

*S1

Relay 1 settings

Off

0

2.0.1

Low setpoint

1

High setpoint

2

High

Note: Main process

“HOLD” active

3

means cond. or resist. Process alarm

0

Alarm

whichever is set in

Proportional duty cycle control **

1

code #1

Proportional frequency control **

2

PI duty cycle **

3

PI pulse freq. **

4

USP contact

5

Calculated value

0

Meas cond/res. value cell 1

1

Cell 1

Meas cond/res. value cell 2

2

Meas temp. value cell 1

3

Meas temp. value cell 2

4

Meas temp. T1-T2

5

41

*S2

Relay 2 settings

Off

0

1.0.1

Low setpoint

1

Low

High setpoint

2

Note: “HOLD” active “HOLD” active

3

relay contact is used Process alarm

0

Alarm

to indicate when the

Proportional duty cycle control

1

measuring mode is

Proportional frequency control

2

interrupted

PI duty cycle **

3

PI pulse freq. **

4

USP contact

5