IM 12D08E02-01E

Parameter setting 5-23

100

50

0

0.3 s

% controller output

Code Display

Function

Function detail

X

Y

Z

Default values

Contacts (continued)
44

*D.TIME

Delay time

Minimum relay switching time

0.2 sec.

*SC.HYS

Process hysteresis

Minimum change of process value

2.0 %

for relay reset after switching

*T.HYST

Hysteresis temp.

Minimum temperature change for relay

1 °C

reset after switching (fig. 5-3)

*C.HYST

Calculation hysteresis

0

45

*RANGE

Proportional range

When proportional control selected

10.0 %

in code 40, 41, 42 or 43

*PER.

Duty cycle period

Pulse control On time + Off time (fig. 5-4)

10 sec.

*FREQ.

Maximum frequency 100% value for frequency control (fig. 5-5)

70 p/m

46

*tI.CNT

Integral time

Integral time for relay controls when

100 sec.

PI is set

47

*EXPIR

Expiry time

Warning of ineffective control action On

1

0

Off

Warning of ineffective control action Off

0

*tE.min

Set expiry time

Set expiring time using >, ^, ENT keys

15 min

48

*SC1

Set control range

Set range for *SC1 (*SC2) for proportional

contact control when *SC1 (*SC2) are not

used on mA1 (mA2)

0%

Set begin scale

Use >,^, ENT keys to set value

*100%

Set end scale

Use >, ^ ENT keys to set value

49

Not used

90%

100

50

0

on

t

t

of f

90%

10%

50%

50%

10

%

Fig. 5-4. Duty cycle control

SC

% of output range

% duty cycle control

Setpoint

Proportional

range

Time

Pulse period

Fig. 5-5. Pulse frequency control

SC

% of output range

Setpoint

Proportional

range

Time

Maximum pulse frequency

50 % pulse frequency

No pulses

Fig. 5-3.

t (sec)

LED off

Setpoint

LED on

Delay time

Delay time

LED off

Cond./Resist.

Hys.