Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 50

Advertising
background image

IM 12D08E02-01E

Parameter setting 5-25

Code Display

Function

Function detail

X

Y

Z

Default values

User interface

50

*RET

Auto return

Auto return to measuring mode Off

0

Auto return to measuring mode On

1

1

On

51

*MODE

Mode setup

Setpoints in maintenance mode Off

0

0

Off

Setpoints in maintenance mode On

1

52

*PASS

Passcode

Maintenance passcode Off

0

0.0.0 Off

Note # = 0 - 9, where Maintenance passcode On

#

0 = no passcode

Commissioning passcode Off

0

Off

1=111, 2=333, 3=777 Commissioning passcode On

#

4=888, 5=123, 6=957 Service passcode Off

0

Off

7=331, 8=546, 9=847 Service passcode On

#

53

*Err. 1. 1(2) Error setting

Polarisation too high

Soft fail

0

Polarisation too high

Hard fail

1

1

Hard

*Err. 5. 1(2)

Shorted measurement

Soft fail

0

Shorted measurement

Hard fail

1

1

Hard

*Err. 6. 1(2)

Open measurement

Soft fail

0

Open measurement

Hard fail

1

1

Hard

*Err. 7. 1(2)

Temperature sensor open Soft fail

0

Temperature sensor open Hard fail

1

1

Hard

*Err. 8. 1(2)

Temp. sensor shorted

Soft fail

0

Temp. sensor shorted

Hard fail

1

1

Hard

*Err. 13. 1(2)

USP <645> limit exceeded Soft fail

0

0

Soft

USP <645> limit exceeded Hard fail

1

*Err.22. 1

Control time-out

Soft fail

0

0

Soft

Control time-out

Hard fail

1

*SOFT

SOFT FAIL CONTACT Pulsing contact

0

0

Pulsing

No Pulsing contact

1

54

*E5.LIM 1(2) E5 limit setting

Maximum conductivity value

25

mS

(Minimum resistivity value)

0.04 kΩ

*E6.LIM 1(2) E6 limit setting

Minimum conductivity value

1

µS

(Maximum resistivity value)

1

55

*%

Display mA in w/w% mA1-range displayed in w/w% off

0

0.0

Off

mA1-range displayed in w/w% on

1

mA2-range displayed in w/w% off

0

Off

mA2-range displayed in w/w% on

1

*%1

Set w/w% for range 1 Press YES to access 0% value adjustm.

*0%

Set 0% output value in w/w%

*100%

Set 100% output value in w/w%

*%2

Set w/w% for range 2 Press YES to access 0% value adjustm.

*0%

Set 0% output value in w/w%

*100%

Set 100% output value in w/w%

56

*DISP

Display resolution

Auto ranging display

0

0

Auto

Display fixed to X.XXX µS/cm or MΩ•cm 1

Display fixed to XX.XX µS/cm or MΩ•cm 2

Display fixed to XXX.X µS/cm or MΩ•cm 3

Display fixed to X.XXX mS/cm or kΩ•cm 4

Display fixed to XX.XX mS/cm or kΩ•cm 5

Display fixed to XXX.X mS/cm or kΩ•cm 6

Display fixed to XXXX mS/cm or kΩ•cm 7

57

*USP

USP setting

Disable the E13 (USP limit passed)

0

0

0.0

Off/Off

X=Cell 1 Y= Cell 2

Enable the E13 (USP limit passed)

1

1

58

*CALC

Calculation setting

Choose the main parameter for display

Ratio (a/b)

0

5

No calc.

Differential (a-b)

1

% passage b/a

2

% rejection (100 (a-b)/a)

3

Deviation (100 X (b-a)/a)

4

No calculation SC1 in display

5

VGB directive 450L, pH calculation

6

When set for resistivity measuring, only

selection 1 and 5 are available