9.4 Settng the Constants Used n Equatons

This section explains the details of setting constants to be used in computing equations.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Math and then press the key.

3.

Press the

key or

SHIFT +

key to select

Constant and then press the

key.

4.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

Press the key to set other channels.
If you are done, press the

ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Select the constant number.

The new setting takes effect.

ESC/?

Set the value.

K01

Setting complete

Set=Math

Math=Constant

No.=K01

K01=1

Explanaton

Constant

Up to 30 constants can be registered. Enter the values as K01 to K30 in the equations.
Set the constants in the range shown below. The maximum number of significant digits is 5.
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29