Measuring accuracy and recording accuracy in case of square root computation

A: Voltage measurement accuracy (digits)

B: Voltage span (digits)

C: Scaling span (digits, upper limit of scaling – lower limit of scaling)

Input Range

Computed Value Accuracy Equation (Digits),

Rounded up to the decimal place

Recording Accuracy

0% or more, less than 1.5%

*

0.1 × C + 2

Left value + 0.3% of

recording span

1.5% or more, less than 6.25% A / B × C × 5 + 2

6.25% or more, less than 25% A / B × C × 2 + 2

25% or more, 100% or less

A / B × C + 2

* Includes case when:

(input voltage – lower limit of the specified voltage range) / used range × 20000 < 256.

If the specified range is, for example, 1–5 V, the lower limit of the specified voltage range is 1.

Reference junction compensation

Internal/External selectable for each channel

Reference junction compensation accuracy

Above 0°C with input terminal temperature balanced (60 minutes after power on)

Type R, S, B, W, WRe:

±1.0°C

Type K, J, E, T, N, L, U: ±0.5°C

Maximum input voltage

±10 VDC (continuous) for ranges of 200 mV or less, TC, RTD, and DI ranges

±60 VDC (continuous) for 2 VDC or higher ranges

Input resistance

Approximately 10 MΩ or more for ranges of 200 mV or less and TC

Approximately 1 MΩ for 2 VDC or higher ranges

Input source resistance

Volt, TC:

2 kΩ or less

RTD input: 10 Ω or less per wire (The resistance of all three wires must be equal).

Bias current

10 nA or less (except when burnout detection function is enabled)

Common mode rejection ratio 120 dB (50/60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)

Normal mode rejection ratio 40 dB or more (50/60 Hz ± 0.1%)

Noise rejection

By the integrating A/D converter: The integration time is 16.7 ms, 20 ms, or 100 ms (dot model)

Low-pass filtering (pen model):

Filter time constant is 2 s, 5 s, or 10 s

Moving average (dot model):

The number of samples is 2 to 16.

12.6 General Specifications