The condenser cooling surface must be kept clean:

-

Remove the rear panel of the station (4 screws).


-

Clean the cooling surface with compressed air and perhaps a soft brush. Be
careful not to bend the fins since this would reduce the air flow and impair
condenser capacity.


-

Replace rear panel.


Check the oil level in the vacuum pump:

-

If the oil level is below the centre of the sight glass, replenish as follows:


-

Remove cap (pos. 32) to replenish vacuum pump.


-

Replenish (slowly) with vacuum oil to the correct level.


- Replace

cap.


For each 75 operating hours:


Replacement of filter drier (pos. 5)

-

Remove front panel (6 screws).


-

Remove filter by loosening the pressure control (two 3/8” flare nuts), and
then fit the replacement filter. Always use a new filter fitted with protective
caps on the connectors.


Replacement of filter drier (pos. 8)

-

Remove rear panel (6 screws).


-

Remove filter by loosening the union nuts and the solenoid valve (pos. 36)
at the filter end. Loosen the nuts slowly and take out the filter.


-

Fit a new filter in the station, making sure to retighten the union nuts.


-

Refit pressure control and solenoid valve.


On replacing filter driers a small quantity of refrigerant escapes - therefore
follow the appropriate safety precautions.