Re-Calibration Only

11

9. The alpha-numeric display reads

HYST. CAL TARGET=XXX

(X represents the preset target value on

display). Set the input pressure from the test standard to the target value on the display. Wait for system
to stabilize and press the

ENTER

key on the display panel.

NOTE: For ABS models, use a vacuum pump to reach the applicable SPAN or HYST target values that are
below local barometric pressure.

10. The alpha-numeric display reads

FILTER 5

. Save the calibration data by opening rocker switch

S1

position

2

. DLR3110 returns to normal mode.

11. Vent system and remove test standard.