DLR3110 Rack Mount Digital Pressure Calibrator Manual

4. The alpha-numeric display reads

SPAN 2 CAL TARGET=X

(X represents the existing target value on

display). Change target value if required. Set the input pressure from the test standard to the target value
on the display. Wait for system to stabilize and press the

ENTER

key on the display panel. See Table 5-7

as required.

5. The alpha-numeric display reads

SPAN 3 CAL TARGET=X

(X represents the existing target value on

display). Change target value if required. Set the input pressure from the test standard to the target value
on the alpha-numeric display. Wait for system to stabilize and press the

ENTER

key on the display panel.

Repeat this step for SPAN target values 4 – 10.

6. The alpha-numeric display reads

HYST.CAL TARGET=XXX

(X represents the preset target value on

display. See Table 5-8.). Set the input pressure from the test standard to the target value on the
alpha-numeric display. Wait for system to stabilize and press the

ENTER

key on the display panel.

NOTE: For absolute models, use a vacuum pump to reach the applicable hysteresis target value if below local
barometric pressure.

7. Save the calibration data by opening rocker switch

S1

position

2

. The DLR3110 returns to normal mode

(see Section 5.3 on page 12).

8. Vent system and remove test standard and/or vacuum pump.

LED Display

Alpha-Numeric Display

Key Pressed

Remarks

CAL

SPAN OK

N/A

Span has been accepted. Proceed to step
4 below. Span may be accepted, but
based on span value, alpha-numeric
display may return with scrolling message
“FULL SCALE WILL NOT BE OBTAINED”.

NO CAL

SPAN CAL

N/A

Span has not been accepted. Repeat step
3 above. If span is not accepted, display
returns with one of the following scrolling
NO CAL messages: “SIGNAL INPUT
UNDER THE MINIMUM ZERO RANGE” or
“INPUT ABOVE CAL INPUT RANGE”.

Table 5-7. Calibration Mode Span Messages

LED Display

Alpha-Numeric Display

Key Pressed

Remarks

*XX

HYST.CAL TARGET=XXX

See step 6 below

To change the target value, see Section 5.7
on page 19.

*XX

FILTR 5

DOWN arrow

To change target value 1 to 10 by 1, 10 to
20 by 2, 20 to 50 by 5, press UP/DOWN
arrow keys.

----

EXIT ?

YES key

Return to main menu display SET-UP/IO ?.
See Section 5.4 on page 13.

NO key

Return to XXX/CAPACITY. See
Section 5.6.1 on page 17.

Table 5-8. Calibration Mode Hysteresis and Filter Target Parameters