DLR3110 Rack Mount Digital Pressure Calibrator Manual

12. Open Pressure Limit Monitor completely, by turning regulator knob clockwise. If you reach between

100-105% of full scale and pressure relief valve was not disturbed regulator has been adjusted properly.

13. Replace regulator knob cap.

7.2.14

Panel Gauge Removal

Tools required:

Phillips screwdriver
7/16" wrench
9/16 " wrench

NOTE: Call CONDEC for replacement Part Numbers.

1. Remove cover from its chassis as described in Section 7.2.1 on page 39, and carefully place on a bench.

2. Disconnect the tubing section that connects to the gauge fitting.

3. Loosen the two thumb-nuts that hold the gauge mounting U-clamp.

4. While gripping the square portion of the gauge port with the 9/16" wrench, remove the Female Tube

Connector (PN 59721) from the gauge.

5. Remove the two thumb-nuts, the mounting U-clamp, and the gauge.

7.2.15

Panel Gauge Installation

Tools required:

Phillips screwdriver
7/16" wrench
9/16" wrench
A/R 1/4" wide Teflon tape (PN 60575)
SNOOP, liquid leak gas detector (PN 64781)

NOTE: Call CONDEC for replacement part numbers.

1. Before installing a new gauge, wrap two layers of new Teflon tape on the port.

2. Install gauge into panel, secure with U-clamp and tighten the two thumb screws.

3. While gripping the square portion of the gauge port with the 9/16" wrench, tighten the Female Tube

Connector on to the gauge.

4. Attach the tubing section that connects to the gauge fitting.

5. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and

check all fittings for leaks.

6. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.

7.2.16

Test Port Quick-Connect Fitting (PN 59762) Removal and Installation - 3 and 100 Full Scale PSI

If there is leakage out of the port, replace the test port fitting.

Tools required:

Phillips screwdriver
5/8" two open end wrenches
9/16" open end wrench
SNOOP, liquid leak gas detector (PN 64781)

1. Remove cover from its chassis as described in Section 7.2.1 on page 39, and carefully set on a bench top.

2. Loosen and remove the tubing end nut from the test port fitting.

3. Grasp the hex nut at the panel face with a 5/8" wrench and using a second wrench, turn the nut on inside

of panel counter-clockwise.

4. Install the new quick-connect fitting (PN 59762) by reversing steps.

5. Using a pressure source set input pressure to approximately 100% of full scale rating of DLR3110 and

check all fittings for leaks.

6. Install panel/chassis assembly in its enclosure as described in Section 7.2.1 on page 39.