DLR3110 Rack Mount Digital Pressure Calibrator Manual

By combining these sensors with microprocessor-based circuitry, an even higher degree of operational accuracy
and precision has been accomplished. For example, computer-generated correction curves for both the
non-linearity and the hysteresis of the sensors improve these characteristics by an order of magnitude or more.

Two micro-metering valves and vernier are provided to control the external nitrogen source while the digital
display indicates the magnitude of the applied test pressure. Also, a push-button switch provides the “zeroing” of
the pressure display (on GAGE models). Over-pressure protection, based on model, is provided via a fully
adjustable pressure regulator which is manually set for each new device being tested.

Figure 1-1 provides an overview of the DLR3110’s function.

Figure 1-1. DLR3110 Flow Diagram

MONITOR

PRESSURE LIMIT

RELIEF

VALVE

INPUT PORT

±50 PSI

3000 PSI

BURST DISK

CONTROL

PRESSURE LIMIT

FILTER

20u

MICROPROCESSOR-BASED

DIGITAL INDICATOR

TRANSDUCER

VENT

VENT/VACUUM

PORT

(QUICK DISCONNECT)

PRESSURE

INCREASE

VERNIER

TEST PORT

20u

FILTER

ADJUST

DLR3110 FLOW DIAGRAM

NOT REQUIRED ON 5K OR 10K MODELS

NOTE: MAX INPUT BASED\P ON MODEL.