Configuration

13

3.0

Configuration

Prior to calibration, the UMC600 must be digitally
configured, or assigned a set of operating parameters.
The three parameters listed in Section 3.1.1 are directly
related to calibration and must be set before proceeding
to calibration mode.

3.1

Digital Configuration

3.1.1

Parameter Overview

Table 3-1 on page 14 lists configuration parameters and
describes their values. The following paragraphs give
the procedure for configuring the UMC600.

3.1.2

Configuration Procedure

1. Unscrew the two screws on the face plate

bracket (Figure 3-1). The bracket drops down,
exposing four program switches on the left side
of the indicator.

Figure 3-1. Accessing the Program Switches

2. Temporarily remove the unit’s flexible black

display overlay by gently pushing down and
lifting the panel up and out at its center to expose
the configuration and calibration instructions
printed on the surface below. The switch
function table defines the appropriate front panel
switch settings for the

CONF

and

CAL

modes.

3. Close switch SW1-2, marked

CONF

(2), by

moving to the up position (see Figure 3-2). A
prompt appears with a parameter number and
data value.

The parameter identifier is a number, 1–14, that
correlates to the

CONFIG

chart on the upper left of the

switch map panel. Selected data represents the value
being entered into the unit configuration data. For
example, 1 100 sets the indicator to 10,000 graduations
(see Table 3-2 on page 16).

When configuration is complete, set SW1-2 down to
return the unit to normal operating mode.

Figure 3-2. Closing Switch 2

Figure 3-3 defines the functional operation of each key
on the front panel of the indicator when the unit is in the
the setup mode.

Figure 3-3. Front Panel Key Functions

NOTE: The TARE RECALL key functions as a previous
screen key in CONFIG mode.

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

ZERO

3 4

2

1

OPEN

Zero

Net/Gross

Tare

Units

Parameter

Select

Parameter Data

Select

Subparameter

Select

Subparameter

Data Select

DEAD LOAD (3)

CONF (2)

CAL (1)

3 4

2

1

OPEN

NORM

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

ZERO

Parameter
Identifier

Selected Data

ZERO

NET

GROSS

TARE

lb/kg

CONV

Advertising