background image

What should I do if …? • 71

English

Error LED flashing

Macro executing

- Wait until macro complete.

Error LED flashing

Power supply too low:

• power supply unit defective

- Replace power supply unit

• printer defective

- Do not continue using printer. Use replacement

device

Error LED flashing

Power supply too high:

• power supply unit defective

- Replace power supply unit

• printer defective

- Use substitute device

"MSG For Children only"
message appears on
the printout

• No reference data are available for the patient age

entered
- Check age entered and correct if necessary

"MSG image not loaded
to printer" message
appears on the printout

• No logo saved on the SD card

- Select a logo and save it on the SD card - see "Select

your own logo (logo)" on page 61

Printing does not start.

• It was not possible to transmit measurement results to

the printer.
- Ensure that the printer is switched on.

• Printer not integrated in wireless network.

- Switch off the scale or height measuring device
- De-energize the printer and its wireless module
- Set up the wireless network again.

• Nearby HF equipment (e.g. mobile phones) is

interfering with reception.
- Make sure that HF equipment is kept at least 1 metre

away from transmitters and receivers in the seca
wireless network.

Incorrect time and date
on printout

• Date and time not set

- Set date and time using a seca 360° wireless scale

or height measuring device (see instruction manual
for the

- seca 360° wireless device used).

• Internal battery of wireless module exhausted.

- Contact seca Service.

Fault