ANALOG GAIN - KFA (Display 4)

The Analog Gain (KFA) is the scaling between the frequency input and the analog output. This is only

valid in Monitor Mode -- in PID Mode, KFA controls the closed loop reaction time (P21).

This should not be confused with the KFR factor

ANALOG GAIN (KFA)
CURRENT=XXXXXXXX

The formula that describes the relation is as follows

:

Frequency when Eng. Unit =Gallons

V * 33400

Imp/gal

X

GPM

KFA =

f =

f

60 sec

where,

V is the desired Output Voltage at frequency f: If 5V - use the GPM realistically achievable by

your system to find f.

- 33400 is a constant

Once the right number is entered press ENT

The default value for KFA=00000167 (1000Hz = 5Volt)

The maximum KFA is 00065000

After the formula has been calculated and the correct number has been programmed in to the KFA, the

Analog Output can be seen by going into the Status Display Mode. The Status Display shows a number

from 0 to 1023 for the Analog Output. This number is the D/A converter input from the

microprocessor. When the EMO-1005 is set for a maximum of 5 Volt DC on the Analog Output (Pin

10) then each step of the converter is equal to:

One step = 5Volt/1024steps = 4.8828mV

Therefore, if 1 Volt is measured on the Analog Output then the Status Display should be showing a

count of 205 in the Analog Out (AO0205) position. (1 Volt = 205steps * 4.8828mV)