Checksum calculation – AW Gear Meters EMO-1005 User Manual

Page 33

Advertising
background image

AW-Lake Company 8809 Industrial Drive, Franksville, WI 53126  web: www.awgearmeters.com

 Tel: 262-884-9800  Fax: 262-884-9810  Email: [email protected]

REV. 1a 05/98 EMO-1005 Manual.DOC

32

Writing with the host computer:

>01J03340F6587XXcr

^....................... start of text always > sign

^^.................... channel number that should be written to

^................... J = writing into channel (K = read)
^^^^........... address to write to
^^......... byte written to 0334
^^...... byte written to 0335
^^... byte written to 0336
^^.. checksum for string 01J03340F6558
^^. carriage return

If write operation was successful the channel answers with an: Acr

The following error messages have been implemented in the EMO-1005 protocol:

N01cr - Undefined command (something else than K/J)

N02cr - Checksum error

N03cr - Non printable ASCII character


There is a booklet published by OPTO 22 on the protocol format.

CHECKSUM CALCULATION

Calculating the previous string variable for the read command:

>03K0F3403XXcr

^^^^^^^^^...... these characters have to be put into the calculation.

Add the values for the ASCII Decimal variables as follows:

HEX

0 + 3 +K + 0 + F + 3 + 4 + 0 + 3

ASCII

48 + 51 + 75 + 48 + 70 + 51 + 52 + 48 + 51 = 494


To calculate the final checksum, convert the decimal number 494 into HEX (494=1EEh) and take the
last two characters (EE) for the checksum.

The final string would look like this:

>03K0F3403EEcr