EMO‐3000

 

Operation and Programming Manual 

 

 

 

37 

      
 

KFR = 6000/(X Pulses/Engineering unit)   (for units/minute) 

Locate the X Pulses per Engineering Unit in the calibration sheets of the flow transmitter. It is best to use 
numbers larger than 100 but less 9000.  

The generic K‐factors for the most common AW‐Gear Meters transmitters are: 

Transmitter 

KFR cc/min 

KFR Gal/min 

KFR Oz/min 

ZHM‐01 

142.0 

.0375 

4.80 

ZHM‐02/1 

720.0 

.1915 

24.51 

ZHM‐02 

1363. 

.360 

46.00 

ZHM‐03 

3448. 

.911 

116.6 

HPM‐15 

720.0 

.1915 

24.51 

HPM‐20 

1363. 

.360 

46.00 

HPM‐30 

3448. 

.911 

116.6 

 

As these are generic K‐factors, consult the calibration sheet supplied with the flow transmitter for the 
most accurate information. 

USED IN: 

 

All modes. 

RELATED VARIABLES: 

 

DP for Rate A KFR (only in OPTO communication). 

HEX ADDRESS: 

 

00E2 HEX  ‐‐‐‐>  same address used in EMO‐2000. 

BYTES: 

 

2 Bytes        DEFAULT: 720 

ASC form COMMAND: 

 

reading = 106; writing = 506; E is with decimal point. N is just the 
numeric value. 

K‐Factor for Total A KFT (including the decimal point)  

The KFT‐factor scales the display to the incoming amount of pulses on channel A, which is on pin 4. Enter 
the decimal point in ASC‐form communication. If the OPTO type communication is used, then you must 
enter the decimal point separately.  

The formula for calculating the KFT value is: 

KFT = 10000/(x Pulses/Engineering unit)   (for pulses) 

The figure X Pulses per Engineering Unit can be found in the calibration sheets of the flow transmitter. It 
is advisable to use numbers bigger than 100 or less 9000. 

The generic K‐factors for the most common AW Gear Meters transmitters are as follows : 