Emo‐3000 – AW Gear Meters EMO-3000 User Manual

Page 39

Advertising
background image

EMO‐3000

 

Operation and Programming Manual 

 

 

 

38 

      
 

Transmitter 

KFT cc 

KFR Gal 

KFR Oz 

ZHM‐01 

236.6 

.0624 

8.00 

ZHM‐02/1 

1200. 

.319 

40.84 

ZHM‐02 

2271. 

.599 

76.66 

ZHM‐03 

5746. 

1.518 

194.0 

HPM‐15 

1200. 

.319 

40.84 

HPM‐20 

2271. 

.599 

76.66 

HPM‐30 

5746. 

1.518 

194.0 

 

As these are generic K‐factors, consult the calibration sheet supplied with the flow transmitter for the 
most accurate information. 

USED IN: 

 

All modes. 

RELATED VARIABLES: 

 

DP for Rate A KFT (only in OPTO communication). 

HEX ADDRESS: 

 

00E4 HEX  ‐‐‐‐>  same address used in EMO‐2000. 

BYTES: 

 

2 Bytes        DEFAULT: 1200 

ASC form COMMAND: 

 

reading = 107; writing = 507; E is with decimal point. N is just the 
numeric value. 

Engineering units for KFR 

The KFR units is stored as a three‐letter string. This means any engineering unit can be constructed if a 
computer hookup is used. 

USED IN: 

 

All modes. 

RELATED VARIABLES: 

 

KFT Engineering units. 

HEX ADDRESS: 

 

0052 HEX  ‐‐‐‐>  same address used in EMO‐2000. 

BYTES: 

 

3 Bytes        DEFAULT: cc/m 

ASC form COMMAND: 

 

reading = 109; writing = 509; E is engineering. 

Engineering units for KFT  

The KFT units is stored as a three‐letter string. This means any engineering unit can be constructed if a 
computer hookup is used. 

USED IN: 

 

All modes. 