Caution – HT instruments HT8000 User Manual

Page 11

Advertising
background image

HT8000

EN - 10

4.3.4. DC current generation

CAUTION

The maximum DC current generated by the instrument is 24mA with DC
voltage of current generator max 24VDC

Fig. 5: DC current generation

1. Press INPUT/OUTPUT key and select the generation mode. The “OUTPUT” symbol is

displayed

2. Press mA/% key and select the desired measurement unit. The symbol “mA” or “mA%”

is displayed

3. Insert the red plug into +LOOP jack and the black plug into V jack
4. Connect the red and black test leads respectively to the positive and negative poles of

the external device which should receive the power supply (see Fig. 5)

5. By means arrow keys

 and  (see Fig. 1 – parts 9, 10, 11 and 12) to set the value of

output generated current. The maximum set value is 24mA DC. For the “mA%” output
consider that -25% = 0mA, 0% = 4mA, 100% = 20mA and 125% = 24mA. The current
value will be displayed

6. For the generation of a negative value of current connect the test leads in the opposite

direction than shown in Fig. 5