Technical specifications – HT instruments HT8000 User Manual

Page 15

Advertising
background image

HT8000

EN - 14

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL FEATURES
Accuracy calculated as [%rdg + (no. dgt x resol.)] at 18°C

28°C (64°F  82°F), <75%RH

DC Voltage (measured and generated)

Range Resolution Accuracy Input impedance

Overload protection

0.01

100.00mV

0.01mV

(0.02rdg +2dgt)

2M

, <100pF

30VDC

0.001

10.000V

0.001V

DC Current (measured and generated)

Range

Resolution

Accuracy

Percentage value Overload protection

0.001

24.000mA

0.001mA

(0.015%rdg+4dgt)

0% = 4mA

100% = 20mA
125% = 24mA

Fast fuse

125mA, 250V

-25.00

125.00%

0.01%

Source mode: 1000

 @ 20mA (battery voltage  6.8V) ; 700 @ 20mA (battery voltage within 5.8V and 6.8V)

Simulating mode: required loop voltage: 24V nominal, 30V maximum, 12V minimum

Loop mode

Range Resolution Accuracy

Overload protection

24V DC

Not necessary

10%

30VDC

6.1.1. Reference

standards

Safety: IEC/EN61010-1
Insulation:

double insulation

Pollution degree:

2

Max height of use:

3000m (9843 ft)

6.1.2. General

data

Mechanical characteristics
Dimensions (L x W x H):

190 x 89 x 42mm, 7 x 4 x 2 inches

Weight (including battery):

350g (12 ounces)

Power supply
Battery type:

1x9V alkaline NEDA1604, 006P, IEC6F22

Low battery indication:

"

" symbol is displayed

Battery life:

4 hours (1604 type), 2hours (006P type)

Auto Power OFF:

after 6 minutes of idleness (disabled)

Display
Specifications:

5 LCD + symbol, decimal point

Over range indication:

O.L” message at display

Negative value indication:

-O.L” message at display for the herewith values:

<-2mV [100mV], <-0.2V [10V], <-1mA [24mA]

Fuses

type FF 125mA/250V, 5x20mm, 0.03kA

6.2. ENVIRONMENT
6.2.1. Environmental

conditions

Reference temperature:

18°C

 28°C ; (64°F  82°F)

Working temperature:

-10 ÷ 55°C ; (14°F ÷ 131°F)

Relative humidity:

<75%RH

Storage temperature:

-40 ÷ 60°C ; (-40°F ÷ 140°F)

This product conforms to the prescriptions of the EMC directive 2004/108/EEC

6.3. ACCESSORIES

 Couple of test leads – Cod. KIT4000A