Caution – HT instruments HT8000 User Manual

Page 12

Advertising
background image

HT8000

EN - 11

4.3.5. Loop current measurement of external transducers

CAUTION

 In this mode the instrument generate a fixed output voltage of 24VDC

which permits to supply an external transducer and perform the
simultaneous current measurement on it

 Maximum input for DC current is 24mA. Do not attempt to take any current

measurement exceeding such limit to avoid the risk of electrical shock and
damages to the instrument

Fig. 6: Loop current measurement of external transducers

1. Switch OFF the circuit under test
2. Press INPUT/OUTPUT key and select the measurement mode. The “INPUT” symbol is

displayed

3. Press mA/% key and select the desired measurement unit. The symbol “mA” or “mA%”

is displayed

4. Insert the red plug into +LOOP jack and the black plug into mA jack
5. Connect red and black plugs to the external transducer respecting the polarity and

current flow (see Fig. 6).

6. Power the circuit under test. The current value will be displayed
7. The message "O.L" means that the current exceeds the measuring limits. For the

generation of a negative value of voltage connect the test leads in the opposite
direction than shown in Fig. 6