7 scpi programming, 1 scpi messages, 2 common commands/queries – KEPCO ATE-DMG SERIES User Manual

Page 56: 3 scpi subsystem command/query structure, Scpi programming -22, Scpi messages -22, Common commands/queries -22, Scpi subsystem command/query structure -22, D 3.7)

Advertising
background image

3-22

ATE-DMG 042314

mode will be, so the unit is never uncontrolled, response is quick and no transients are possible.
Recommended programming techniques are:

1. Minimize programmed mode (voltage or current) changes. Unless absolutely required by the

test parameters, allow the power supply to automatically switch modes as determined by the
load. This will improve response time and reduce undesirable transients. For those power
supplies that employ relays (Kepco's MBT with “R” option, MAT and MST) this will also
increase the life of the relay.

2. Once the mode (voltage or current) is programmed, program the active parameter to zero

and the complementary limit parameter to the maximum anticipated for application. Then
program only the active parameter. The active parameter is the parameter that controls the
output, e.g., voltage controls the output in voltage mode.

3. Never program both the active and complementary limit parameter to zero. This can result in

long response times. Set the active parameter to zero and the complementary limit parame-
ter to a minimum, e.g., 10% of maximum, to ensure that the active mode is defined.

3.7

SCPI PROGRAMMING

SCPI (Standard Commands for Programmable Instruments) is a programming language con-
forming to the protocols and standards established by IEEE 488.2 (reference document
ANSI/IEEE Std 488.2, IEEE Standard Codes, Formats, Protocols, and Common Commands). SCPI com-
mands are sent to the ATE-DMG Power Supply as ASCII output strings within the selected pro-
gramming language (PASCAL, BASIC, etc.) in accordance with the manufacturer’s
requirements for the particular GPIB controller card used.

Different programming languages (e.g., BASIC, C, PASCAL, etc.) have different ways of repre-
senting data that is to be put on the IEEE 488 bus. It is up to the programmer to determine how
to output the character sequence required for the programming language used. Address infor-
mation (GPIB address) must be included before the command sequence. (See PAR.3.6.3 to
establish the ATE-DMG Power Supply GPIB address.)

3.7.1

SCPI MESSAGES

There are two kinds of SCPI messages: program messages from controller to power supply,
and response messages from the power supply to the controller. Program messages consist of
one or more properly formatted commands/queries and instruct the power supply to perform an
action; the controller may send a program message at any time. Response messages consist of
formatted data; the data can contain information regarding operating parameters, power supply
state, status, or error conditions.

3.7.2

COMMON COMMANDS/QUERIES

Common commands and queries are defined by the IEEE 488.2 standard to perform overall
power supply functions (such as identification, status, or synchronization) unrelated to specific
power supply operation (such as setting voltage/current). Common commands and queries are
preceded by an asterisk (*) and are defined and explained in Appendix A (see Table 4-4). Refer
also to syntax considerations (PARs 3.4.3 through 3.4.6).

3.7.3

SCPI SUBSYSTEM COMMAND/QUERY STRUCTURE

Subsystem commands/queries are related to specific power supply functions (such as setting
output voltage, current limit, etc.) Figure 3-8 is a tree diagram illustrating the structure of SCPI