Table 3-1. lcd messages, 4 keypad functions, Table 3-2. key functions – KEPCO BHK-MG 40W (Half Rack) Series User Manual

Page 46: Keypad functions -2, Lcd messages -2, Key functions -2, Able 3-2 for, D 3-2 expla

Advertising
background image

3-2

BHK-1/2-MG 121313

3.2.1.4

KEYPAD FUNCTIONS

Keypad functions are listed in Table 3-2. Six keys have dual functions, depending on whether
the power supply is in command entry status (waiting for a command to be entered), or data
entry status (waiting for a number to be entered). Command entry status is indicated by a blink-
ing colon

(:_:_:)

and data entry status is indicated by a blinking equal sign

(=_=_=)

.

NOTE: Keys with dual functions are labeled with both a command and a number. The com-

mand label is referred to when the unit is in

(:_:_:)

command entry status; the num-

ber is referred to when the unit is in

(=_=_=)

data entry status.

TABLE 3-1. LCD MESSAGES

LOCATION

MESSAGE

DESCRIPTION

Top left

Loc/Rem

Digital control status: either Remote or Local

Top middle

SINK

Shows that the unit is in sink mode and is absorbing energy either
from the load or the internal output capacitor. The sink current is
model-dependent and appears on the second line of the display pre-
ceded by a minus (–) sign.

top right

OFF/CV/CC

Off/Constant voltage mode/constant current mode

Bottom

left:

in command entry

n.nV

Actual output voltage

In data entry

(parameter)

e.g. OVset if OV SET key was pressed.

Bottom middle:

(:_:_:)

Command entry status

Top or Bottom middle:

(=_=_=)

Data entry status

Bottom

right:

in command entry

n.n mA

Actual output current.

In data entry

n.n V or n.n mA

Present programmed value of parameter, replaced by data entered.

(–)

Minus sign Indicates current absorbed either from the load or dis-
charge of the internal output capacitor.

TABLE 3-2. KEY FUNCTIONS

KEY

POWER SUPPLY

STATUS ACTIVE

DESCRIPTION

REFERENCE

PARAGRAPH

OUTPUT

ON/OFF

Command Entry

If top line of LCD reads OFF, press to enable the output. If output is on,
press to disable the output controlled.

3.2.7

V SET

Command Entry

Press to set output voltage. After V SET is pressed, previous setting is dis-
played. Data entry required to enter new value of output voltage; press
ENTER to accept displayed value.

3.2.9

I SET

Command Entry

Press to set output current. After I SET is pressed, previous setting is dis-
played. Data entry required to enter new value of output current; press
ENTER to accept displayed value.

3.2.9

LOCAL

Command Entry

If the power supply is in remote mode, keypad is disabled except for
LOCAL key. Press to enable keypad. If LCD reads KEYPAD LOCKED, the
LOCAL key is also disabled and can only be unlocked by remote opera-
tion.

3.2.4, B.82

RESET

Command Entry

Press to restore the power on default values: CV mode, output voltage

= 0, output current = value chosen from the POWER-UP OR RESET

CURRENT submenu, overvoltage and overcurrent to 1.1 x nominal value.

3.2.8