KEPCO BHK-MG 40W (Half Rack) Series User Manual

Page 47

Advertising
background image

BHK-1/2-MG 121313

3-3

MENU

Command Entry

Press to enter Menu commands: press repeatedly to scroll through Menu
functions: (1) set LCD contrast, (2) GPIB address, (3) Baud Rate. (4) Loop
Back Test (5) DCL control, (6) Power-Up Digital DC Output on/off, (7)
speaker on/off, (8) change calibration password, (9) Go to Previous cali-
bration, (10) Go to Factory Calibration, (11) view firmware version number,
(12) set Power-up and RESET current, (13) set maximum voltage, (14)
set maximum current. Press ENTER, RESET or CLEAR to exit Menu.

(1) 3.2.5, (2) 3.3.3
(3) 3.5.1, (4) 3.5.4
(5) 3.3.2,
(6) 3.2.7.4
(7,)3.2.6, (8) 4.6,
(9) 4.7, (10) 4.8
(11) 3.2.14
(10) 3.2.8.1
(11, 12) 3.2.11

OV SET

7

Command Entry

Press to set overvoltage protection value. Data entry required to enter the
overvoltage protection value; press ENTER to accept displayed value.

3.2.10

Data Entry

Press to enter number 7.

3.2.1.2

OC SET

8

Command Entry

Press to set overcurrent protection value. Data entry required to enter new
overcurrent protection value; press ENTER to accept displayed value.

3.2.10

Data Entry

Press to enter number 8.

3.2.1.2

CALIB

9

Command Entry

Press to enter Calibration status. Requires password entry; instructions
appear on LCD.

4.3

Data Entry

Press to enter number 9.

3.2.1.2

STORE

Command Entry

Press to store present values of output voltage and current and overvolt-
age and overcurrent protection. Data entry required to select memory
location where values are to be stored

3.2.12

EDIT

PROG

Command Entry

Press to select the starting memory location to be edited. Then use

or

to view or modify the parameters of a specific memory location or to

create a new program.

3.2.15.1

STEP

4

Command Entry

Press to select starting address of program to be executed one step at a
time. Address is updated each time STEP is pressed.

3.2.15.3

Data Entry

Press to enter number 4.

3.2.1.2

TIME

5

Command Entry

Press to edit time value for specific memory location. Select memory loca-
tion (data entry), then enter time value, 0.01 to 655.35 seconds (data
entry).

3.2.15.1.1

Data Entry

Press to enter number 5.

3.2.1.2

RUN

6

Command Entry

Press to run a program. Requires data entry to select starting location.
Press ENTER to accept displayed value.

3.2.15.2

Data Entry

Press to enter number 6.

3.2.1.2

RECALL

Command Entry

Press to recall previously stored values of output voltage and current, and
overvoltage and overcurrent protection. Data entry required to select
memory location containing values to be recalled; press ENTER to accept
displayed value.

3.2.13

CLEAR

Data Entry

Press to exit Data Entry status; any numbers entered are lost. Restores
Command Entry status

3.2.9

1

Data Entry

Press to enter number 1.

3.2.1.2

2

Data Entry

Press to enter number 2.

3.2.1.2

3

Data Entry

Press to enter number 3.

3.2.1.2

ENTER

Data Entry

Press to accept data entered and return to Command Entry status.

3.2.9

TABLE 3-2. KEY FUNCTIONS (CONTINUED)

KEY

POWER SUPPLY

STATUS ACTIVE

DESCRIPTION

REFERENCE

PARAGRAPH