KEPCO BHK-MG 40W (Half Rack) Series User Manual
Page 97
BHK-1/2-MG 121313
3-53
c. Program overcurrent value of unit A (IocsetA) to 1.1 x current limit for power assembly
(step 3a).
d. Program overvoltage value of unit A (VovsetA) to 1.1 x VsetA (step 3b).
4. Program unit B as follows:.
a. Program IsetB to approximately 1.01 to 1.02 x IsetA (step 3a). This maximum current
limit is never used except in the unlikely event that the current limit of Unit A is not work-
ing.
b. Program VsetB to (Eo – VsetA) (VsetA is set in step 3b).
c. Program overcurrent value of unit B (IocsetB) to 1.1 x IsetB (step 4a).
d. Program overvoltage value of unit B (VovsetB) to 1.1 x VsetB (step 4b).
5. To apply power to the load:
a. If load switch used, close the switch.
b. If load switch not used, press POWER ON/OFF key (once) on unit B, then unit A.
6. To disable power to the load:
a. If load switch used, open the switch.
b. If load switch not used, press POWER ON/OFF key once: first on unit A, then on unit B.
c. Alternative: Turn both units off by setting POWER switch to off (down) position, first on
unit A, then on unit B.