Use the following troubleshooting guide for hints on problems with
instrument operation.

8.Before Sending for Repair

Transformer jaws buzz

when measuring large

current.

───

This is not a failure.

Condition

Display blanks after power-

on.

Battery is improperly

installed.

Battery is exhausted.

Install the battery correctly.

Replace the battery.

Possible Cause

Remedy

Display reading remains

frozen.

Data hold button is pressed

in.

Peak hold switch is at ON

position.

Release data hold button.

Set peak hold switch to

OFF position.

The lowest digit of reading

is unstable.

───

This is not a failure.

The instrument is highly

accurate, so it senses

slight variations in the

current under test.

Output cord (Model 7073)

dose not output voltage.

The cord is open is circuit.

Check the cord for an open

circuit.