― 3 ―

CAUTION

●Make sure that the function selector switch is set to an

appropriate position before making measurement.

●Be sure to set the function selector switch to the “OFF”

position after use. When the instrument will not be use for a long
period of time, place it in storage after removing the battery.
This is to avoid damage to the instrument by possible leakage
from the battery.

●Do not expose the instrument to the direct sun, extreme

temperatures or dew fall.

●Placing the instrument in temperatures of 50℃ or higher can

cause the instrument's case to deform and result in operation
failures.

●Never apply voltage to the OUTPUT terminal. the terminal is not

electrically isolated from the internal circuits of the instrument.

Measurement categories (Over-voltage categories)
To ensure safe operation of measuring instruments, IEC61010
establishes safety standards for various electrical environments,
categorized as CAT

to CAT, and called measurement

categories. Higher-numbered categories correspond to electrical
environments with greater momentary energy, so a measuring
instrument designed for CAT

environments can endure greater

momentary energy than one designed for CAT

.

CAT

: Secondary electrical circuits connected to an AC electrical

outlet through a transformer or similar device.

CAT

: Primary electrical circuits of equipment connected to an

AC electrical outlet by a power cord.

CAT

: Primary electrical circuits of the equipment connected

directly to the distribution panel, and feeders from the
distribution panel to outlets.

CAT

:The circuit from the service drop to the service entrance,

and to the power meter and primary over-current
protection device (distribution panel).

Incoming wire

CAT.IV

Interior wiring

Transformer

CAT.I

CAT.II

CAT.III

Socket