Measuring Ranges and Accuracy

●AC Current

3.Specifications

※Refer to Figure 1 for frequency characteristics.

●Effect of External Stray Magnetic Field

10mA AC max. in proximity to a 10mm-dia conductor carrying 100A
AC.

●Effect of Residual Current

10mA AC max. when clamping on two 10mm-dia conductors, each
carrying supply or return 100A AC current.

Ranges

200mA

2A

20A

200A

1000A

0∼199.9mA

0∼1.999A

0∼19.99A

0∼199.9A

0∼500A

501∼1000A

Accuracy

Frequency Range Response

50/60Hz

WIDE

±1.0%rdg±2dgt

(50/60Hz)

±3.0%rdg±2dgt

(40∼1kHz)

±1.5%rdg±2dgt

(50/60Hz)

±3.5%rdg±2dgt

(40∼1kHz)

±1.5%rdg±2dgt

(50/60Hz)

±3.5%rdg±2dgt

(40∼1kHz)

±5%rdg

(50/60Hz)

±10%rdg

(40∼1kHz)

±1.5%rdg±2dgt

±2.0%rdg±2dgt

±2.0%rdg±2dgt

±5.5%rdg

10min.

Time Limit for
Measurements

Continuous