KYORITSU 3023 User Manual

Page 15

Advertising
background image

(6) Set the Range selector switch to the OFF position, and disconnect

the probes from the instrument.

DANGER

¡ Never touch the circuit under test immediately after testing.

Capacitance stored in the circuit may cause electric shock.

Leave the test leads connected to the circuit until the LIVE LED

and LIVE circuit warning stop flickering.

6-3

Continuous measurement

For continuous measurement, a lock-down feature is incorporated on

the Test button. Pressing and turning clockwise locks the button in

the operating position, the button is released by turning it

counterclockwise.

DANGER

¡While the Test button is locked down, a high voltage is present at

the tip of a probe. Attention should be paid to avoid possible shock

hazard.

6-4

Output voltage characteristics

This instrument conforms to IEC61557. This standard defines that the
nominal current shall be at least 1mA, and the lower limit of the

insulation resistance maintaining the nominal voltage at the

measurement terminal. (See the graph below.)

This value is calculated by dividing the nominal voltage by nominal

current. i.e., in case that the nominal voltage is 500V, the lower limit

of the insulation resistance is found as follows.

Divide 500V by 1mA equals 0.5MΩ.

That is, insulation resistance of 0.5MΩ or more is required to provide

the nominal voltage to the instrument.

― 13 ―

Nominal voltage

1MΩ

1000V

0.5MΩ

500V

0.25MΩ

250V

0.125MΩ

125V

0.100MΩ

100V

50kΩ

50V

Lower limit of ins-

ulation resistance

to supply nominal

current 1mA

Advertising
This manual is related to the following products: