A : intrinsic error(%) b : operating error(%) e, Variation due to changing the position(%) e, Variation due to changing the supply voltage(%) e – KYORITSU 3023 User Manual

Page 7

Advertising
background image

― 5 ―

Nominal voltage
Auto Range
Open circuit
voltage
Short circuit
current
Nominal test
current
1

st

effective

measurement
range
Center scale
range
Accuracy

2

nd

effective

measurement
range

Accuracy
Accuracy at 0

KEW3023

1000V

0∼4/40/400/2000MΩ

Nominal test voltage 0%∼+20%

Within 1.5mA

1MΩ

1mA 0%∼+20%

0.100∼

1000MΩ

50MΩ

0∼

0.099MΩ

1001∼

2000MΩ

200.1∼

2000MΩ

40.0∼

2000MΩ

20.01∼

200.0MΩ

0∼

0.099MΩ

0∼

0.099MΩ

0∼

0.099MΩ

±(2%rdg+6dgt)

±(5%rdg+6dgt)

Within 6dgt

50MΩ

5MΩ

5MΩ

0.100∼

200.0MΩ

0.100∼

40.0MΩ

0.100∼

20.0MΩ

0.5MΩ

0.25MΩ

0.1MΩ

500V

250V

100V

0∼4/40/200MΩ

<Operating error>

Operating error (B) is an error obtained under the nominal operating
conditions, and calculated with the intrinsic error (A),
which is an error of the instrument used, and the error (En) due to
variations. According to IEC61557, the maximum operating error
should be within +/-30%.

B =︱A︱+ 1.15 × √

( E

1

2

+ E

2

2

+ E

3

2

A : Intrinsic error(%)
B : Operating error(%)

1

: Variation due to changing the position(%)

2

: Variation due to changing the Supply voltage(%)

3

: Variation due to changing the temperature(%)

Nominal operating conditions
Ambient temperature

: 0∼40℃

Relative humidity

: 90% or less

Position

: Horizontal∼±90゜

Battery voltage

: within the battery effective range

Advertising