7.functions keys – KYORITSU 3023 User Manual

Page 18

Advertising
background image

7-1 Comparator function

Set a reference resistance value at the insulation resistance range.
The measured value is compared to the pre-set reference value, and
the mark of

or

is displayed on the LCD and buzzer

Settable reference value:

0.1/0.2/0.25/0.4/0.5/1/2/3/5/10/20/30/50/100/Any : Unit [MΩ]
The upper limit of

"

Any

"

varies depending on output voltage.

50/100/125V: 0.000 to 199.9: Unit [MΩ]
250/500/1000V: 0.000 to 999.9 : Unit [MΩ]

This function can be set at each range.

Setting procedure

(1) Set the Range selector switch to any desired insulation resistance

range with the

key

pressed down.

In this case, there is no need to press the Safety

key

to set the

range selector switch to 500V or 1000V range.

(2) Press the UP or DOWN

key

to select

or

and then press

the

key

.

Select

to sound the when the measured value exceeds the

pre-set value.
Select

to sound the when the measured value is under the

pre-set value.

(3) The number displayed at the lower left on the LCD starts

flickering. Press the UP or DOWN

key

to select the reference

value, and press the

key

.

(4) The message

and the set value are displayed on the LCD

once the setting completes.

(5) Press the

key to release this function.

Setting for Any

(1) The decimal point displayed at the lower left on the LCD starts

flickering when "Any" is selected. Press the UP or DOWN

key

to

select the desired decimal position, and then press the

key

.

(2) The four-digit start flickering from the left. Press the UP or DOWN

key

and set it to the desired value, and press the

key

.

e.g.: in case 0.5M

Ω

is set.

― 16 ―

7.Functions keys

Fig. 12. Change of sub-indication