3 general specification – KYORITSU 6016 User Manual

Page 16

Advertising
background image

13

Earth Resistance (EN61557-5)

Operating range compliant with

EN61557-5 operating error

Maximum percentage operating error

5.00~1999Ω

±30%

The influencing variations used for calculating the operating error are denoted as
follows;
Temperature : 0 ℃ and 35 ℃
Series interference voltage : 3V
Resistance of the probes and auxiliary
earth electrode resistance : 100 x RA, 50kΩ or less
Supply voltage : 8V to 13.8V

5.3 General specification

Instrument dimensions 235 X 136 X 114mm
Instrument weight:-

1350g (including batteries.)

Reference conditions

Specifications are based on the following
conditions except where otherwise stated:-
1. Ambient temperature: 23±5° C:
2. Relative humidity 45% to 75%
3. Position: horizontal
4. AC power source 230V, 50Hz
5. DC power source: 12.0 V, ripple content 1% or less
6. Altitude up to 2000m, Indoor use

Battery type

Eight LR6 or R6 batteries.

Operating temperature
and humidity.

0 to +40℃ , relative humidity 80% or less, no condensation

Storage temperature
and humidity

-20 to +60 ℃ , relative humidity 75% or less, no condensation.

Display

Dot Matrix LCD 160(W) X 240(H) pixels.

Overload protection

The continuity test circuit is protected by a 0.5A/600V
fast acting (HRC) ceramic fuse mounted in the battery
compartment, where a spare fuse is also stored.
The insulation resistance test circuit is protected by a
resistor against 1000 V AC for 10 seconds.

Advertising