(3) Connect Test Lead M-7129 to Rpe terminal and clip the metal part of DUT

with alligator clip.

(4) Press the

START

switch when above preparation is completed.

(5) First, measure the protective conductor resistance.

When it passes(1

Ω

or less),

lights up in green and display the

measured result on the LCD.

(6) Then the instrument automatically performs insulation resistance

measurement. When it passes(1M

Ω

or more),

lights up in red

and the measurement is completed. The measured values of protective

conductor resistance and insulation resistance, which are passed the test,

will be alternately displayed on the LCD.

(7) When it fails,

or

will light up in red, and "no" is

displayed on the LCD.

The measured value, which is failed the test, and the message "no" will be

alternately displayed on the LCD.

CAUTION

Follow the procedure described in 5.2-3 and do NULL setting before a
measurement.

Alligator clip must make good contact with the enclosure of the DUT.

When the terminal is open or the resistance value exceeds measuring
range, "OL" mark (over range display) appears on the LCD.

Do not touch the device under test while testing is in progress.
Since a high voltage of 500V, user may get electrical shock.

Insulation

Insulation

Earth

Continuity

Earth

Continuity