19

When the fuse blows during use, please replace with new one according to
below procedure.

(1) Use a flat head screwdriver and turn it about 45

°

to left and remove the

fuse cap and fuse.

(2) Remove the fuse from the fuse cap and replace it with new one.
(3) Install the fuse cap and fuse again. At that point, the screwdriver groove

shall be at about 45

°

turned to left from the initial position. Use the flat

head screwdriver and turn it to right.

(The screwdriver groove will stop at the horizontal position.)

WARNING

Be sure to remove mains cord from the instrument before replacing fuse.

The fuse that user can replace is this fuse only. Never attempt to perform
the other repairing.

CAUTION

Please use the specified fuse

(Fast acting type ceramic fuse: 600V/500mA -

Φ

6.3

×

32mm).

For the specified fuse, purchase it by yourself or order it from our agency.