KYORITSU 6201 User Manual

Page 8

Advertising
background image

6

1

)Test socket

Insert the mains plug of DUT to this socket for the polarity test of protective
conductor resistance, insulation resistance and Extension leads.

2

Terminal block

Connect the attached mains cord and Test Leads to this terminal block.

3

Terminal for mains cord

This terminal is connected to a mains supply via M7123.

4

Terminal for Extension leads adaptor

It corresponds to L, N, E of test socket, and the extension leads adaptor(M-
7140) connected with the cord reel to be plugged to it.

5

Rpe terminal

Connect the Test Lead with alligator clip(M7129)(13) to this terminal for the
measurement of protective conductor resistance, and clip the metal parts of
DUT with the alligator clip.

6

LCD

Measured value is displayed.

7

LED for test result

When the value of protective conductor resistance and insulation resistance
exceeds the limit dictated by applicable standards, LED lights up in red.
When it is within the limit, LED lights up in green.

8

Start switch

A measurement starts by pressing this switch.

9

NULL/250V-500V switch

Earthed appliance test measurementIt is used in order to push the
NULL button before protection earth resistance measurement and to
cancel the resistance of a test leads.

Double insulation test measurement
The test voltage of insulation resistance is changed to 500V and 250V.

10

Range switch

Select a range with this switch.

Advertising