9.2

Calibration (

CAL)

+

Before calibration, allow the unit to warm up (see Technical Data on page 54).
After delivery, a % value is displayed as a measured value (related to the adjusted measuring range).
For this reason, the Uniflex SG 45 must be adapted to its measuring task accordingly. To adapt the unit, realize the cal -
ibration correctly.

+

If necessary, the required display unit (

D.Unt) must be selected accordingly.

g

Load cells: the dead load (raw value) provides a signal. Set

InL.1 = 0 to set this signal to zero (span start).

g

With pressure sensors, the sensor should be de-pressurized, or pressure setting for span start:

Calibration 1. step: Define span start

Display

:

InL.1 = OFF

Press keys

p/q

to display the current measurement value.

InL.1 = % measured value (display changes between InL.1 and instantaneous mea-
sured value). With

ô followed by storage of

InL.1 = instantaneous measured value,

related display value OuL.1 = 0

When the enter key ô is pressed immediately, without pressing

p/q the increment / decre-

ment keys, InL.1 = 0 %, OuL.1 = 0 is taken over as standard value.

Calibration 2. step: Define calibration value

g

Load cell

melt pressure sensors:

Load load cells with a defined
load (reference weight; e.g.
75kg)

Enter pressure for span end or a defined known calibration
value.
For sensors with integrated calibration resistor an automatic
switching of the calibration resistor is done.
For the first calibration

SEtit is recommended to configure

CAL.M = 1 , -> S. 47

InH.1 = OFF

Press keys

p/q

to display the instantaneous measured value.

InH.1 = swaps with the

display of the measured value
With the enter key

ô

the instantaneous mesured value is taken over to

InH.1

When the enter key

ô is pressed immediately, without pressing p/q the increment / de-

crement keys

,

InH.1 = 100 %, OuH.1 = Out.1 is taken over.

(The calibrating resistor is switched off automatically)

Calibration 3. step: Calculation of the upper calibrationpoint

OuH.1

With

OuH.1

the upper calibrating point is realized using the

p/q keys according to

the calculated value

Calculation for calibrating resistor (e.g. 80% )
Example: The calibrating resistor has a range from 0..400
Bar; the calibrating resistor simulates 80% of the endvalue,
80% of 400 results 320.

OuH.1is adjusted to 320.

Press

ô

to store OuH.1 = adjusted upper calibration point

g

DonE

donE (duration 1s) storage of InL.1, OuL.1, InH.1, OuH.1 takes place. Additionally, any
previously realized zero setting (function 1) is deleted.

+

A tare function activated before calibration must be de-activated. Normally, it is not purposeful to keep the old tare
condition. If necessary, the tare function can be re-activated.
The zero offset is deleted automatically.

+

The InL.1 and InH.1 values stored during calibration are stored with full resolution.

Installation and calibration

48