Alicat P Series Pressure Gauge User Manual

Page 16

Advertising
background image

16

MAIN

This mode defaults on power up, with

fl ow as the primary displayed parameter.
The following parameters are displayed in

the Main mode.
Line Pressure shows the pressure in the

units specifi ed at ti me of order.
Tare: Pushing the TARE P butt on tares

the pressure gauge and provides it with a

reference point for zero pressure.
This is an important step in obtaining

accurate measurements. It is best to

zero the pressure gauge each ti me it is

powered up. If the pressure reading varies signifi cantly from zero aft er an initi al

tare, give the unit a minute or so to warm up and re-zero it.
If in doubt about whether the pressure is zero, remove the gauge from the

line and open both ports to atmosphere before entering the Tare command.

For liquid pressure devices, all liquid must be drained from the gauge and any

plumbing between the gauge and the atmosphere.
If the unit reads signifi cantly diff erent than zero when it is exposed to

atmospheric pressure, it is a good indicati on that it was given a false tare.

MENU: Pressing MENU switches the screen to the Select Menu display.

MENU

TARE P

PRESS

PSIG

PSIG

+13.60

+

.

Flashing Error Message: An error message (POV = pressure

overrange) fl ashes when pressure exceeds the range of the sensor.

When any item fl ashes, the pressure measurement is not accurate.

Reducing the pressure to within specifi ed limits will return the unit to normal

operati on and accuracy.

If the unit does return to normal operati on contact Alicat.

Do Not Att empt To Tare Absolute Pressure (psia) Instruments!