Alicat P Series Pressure Gauge User Manual

Page 26

Advertising
background image

26

LOOP

VAR

OFF

AUTO

PID

MAIN

SETPT

SOURCE

CONTROL SETUP (conti nued)

Tareing (or zeroing) a gauge pressure or

diff erenti al pressure controller provides it

with a reference point for zero pressure.
OFF AUTO / ON AUTO—this feature allows

you to tare the controller.
The controller must be left in the default

OFF AUTO mode except when actually

tareing the controller as explained below.
It is, however, very important to perform

this adjustment only when you are

certain that the process ports are open to

atmosphere and that there is No Flow!

For liquid pressure devices, all liquid must be drained from the system.
To correctly tare a gauge pressure or diff erenti al pressure controller:

1. Be sure the unit is in the OFF AUTO default setti ng.
2. Disconnect all plumbing and make sure there is No Flow.
3. Push the butt on below OFF AUTO once so that the display reads ON AUTO.
4. Enter a Set-Point of ZERO. A zero set-point results in the closing of the valve

and a known “no fl ow” conditi on.
5. Wait at least 30 seconds.
6. Push the butt on below ON AUTO once so that the display reads OFF AUTO.
7. Reconnect the plumbing.

If the unit reads signifi cantly diff erent than zero, when removed from

the line and open, it is a good indicati on that it was given a false zero.

If your pressure controller was ordered with the opti onal “Tare-P”, you

may tare it following the instructi ons for tareing a pressure gauge on

page 16

Do Not Att empt To Tare Absolute Pressure (psia) Instruments!