MAINTENANCE AND RECALIBRATION

General: P, PC, PCR and PCD-Series Pressure Gauges and Controllers require

minimal maintenance. They have no moving parts. The single most important thing

that affects the life and accuracy of these devices is the quality of the gas being

measured. The instruments are designed to measure CLEAN, DRY, NON-CORROSIVE

gases. If your application requires an aggressive or corrosive gas, please consider

Alicat’s PS, PCS, PCRS and PCDS Series instruments (see page 60).
Recalibration: The recommended period for recalibration is once every year. A

label located on the back of the controller lists the most recent calibration date.

The controller should be returned to the factory for recalibration within one

year from the listed date. Before calling to schedule a recalibration, please note

the serial number on the back of the meter. The Serial Number, Model Number,

and Date of Manufacture are also available on the Model Info display (page 20).
Cleaning: P, PC, PCR and PCD-Series Pressure Gauges and Controllers require

no periodic cleaning. If necessary, the outside of the controller can be cleaned

with a soft dry cloth. Avoid excess moisture or solvents.
For repair, recalibration or recycling of this product, contact:

Alicat Scientific, Inc.

7641 N Business Park Drive

Tucson, Arizona 85743

USA

Ph. 520-290-6060

Fax 520-290-0109

email: [email protected]

Web site: www.alicat.com

Pressure Conversion Table

psi

1.00 =

51.7150 mmHg

psi

1.00 =

2.0360 inHg

psi

1.00 =

27.7080 inH20

psi

1.00 =

68.9480 mbar

psi 100.00 =

6.8046 atm

psi

1.00 =

51.7150 torr

psi

1.00 =

6.8948 kPa

mmHg 100.00 =

3.9370 inHg

mmHg

100 =

1.9337 psi

inHg

100 =

49.1159 psi

inH2O

100 =

3.6091 psi

mbar

100 =

1.4504 psi

atm

1 =

14.6959 psi

torr

100 =

1.9337 psi

kPa

100 =

14.5037 psi

inHg

1 =

25.4000 mmHg